NSN 6525-01-488-4722 · NIIN 014884722 · FSC 6525
NO DASHES 6525014884722
In Stock
X-RAY APPARATUS,RADIOGRAPHIC,MED
Imaging Equipment and Supplies: Medical, Dental, Veterinary
Technical characteristics
3 rows
Straight from the federal item record. Entries the record leaves empty are shown empty — nothing is filled in on its behalf.
| MRC | Requirement | Value |
|---|---|---|
| CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | X-RAY APPARATUS RADIOGRAPHIC MED LIGHTWEIGHT PORTABLE |
| FEAT | SPECIAL FEATURES | LIGHTWEIGHT, PORTABLE/MOBILE DURABLE DIAGNOSTIC X-RAY SYSTEM;PACKED IN A REUSABLE CONTAINER; 110VAC,3 PRONG, 20 AMP; COTS ITEM; DEPT OF HEALTH & HUMAN SERVICE CERTIFIED; CAPABLE OF LONG TERM STORAGE IN AND TRANSPORT THRU CLIMATIC DESIGN TYPES; OPERATION TEMP AT 0 TO 37.8 DEG C; 95% RELATIVE HUMIDITY, NON-CONDENSING;LOWEST SELECTION FOR X-RAY TUBE NOT LESS THAN 25MA; CAPABLE OF PROVIDNG 200MAS EXPOSURES AT ALL KV SETTINGS; EXPOSURE DURATION: OPERATOR SELECTABLE; TIME NOT AFFECTED BY LINE FREQUENCY; CHILDREN TO LARGE ADULT; UNIT IS INDENTED TO BE USED WITH A COMPUTED RADIOGRAPHY PLATE AS THE IMAGE RECEPTOR; X-RAY TUBE HAS STATIONARY ANODE W/ 1.0MM NOMINAL FOCAL SPOT OR SMALLER; TARGET ANGLE PROVIDES FULL COVERAGE OF 14 X 17 IN. FILM AT 40" SID; PNEUMATIC TIRES FOR EASY TRANSPORT; CONTROL UNIT CONTAINS EXPOSURE SWITCH W/6 FT CORD; FOR USE IN FIELD MILITARY MEDICAL FACILITIES W/CEILING HEIGHT OF 75 IN. |
| SUPP | SUPPLEMENTARY FEATURES | CAPABLE OF BEING LIFTED, MOVED AND ASSEMBLED BY TWO PEOPLE WHEN PACKED; TOTAL POWER REQUIREMENTS OF OPERATING UNIT NO MORE THAN 2400 WATTS; VOLTAGE LINE DROP REGULATED TO NO LESS THAN OR EQUAL TO 10%; KV, MA AND EXPOSURE TIMES SETTINGS NOT AFFECTED BY VOLTAGE FLUCTUATIONS |
Cross-references
2 manufacturer part numbers on file for this NSN — each shown against the CAGE code it belongs to.
| Manufacturer part number | CAGE |
|---|---|
| MODEL 6258 | 64616 |
| 01-0360 | 89875 |