NSN 6632-01-234-6794 · NIIN 012346794 · FSC 6632
NO DASHES 6632012346794
In Stock
ANALYZER,CLINICAL CHEMISTRY
Medical Laboratory Analysis Instruments
Technical characteristics
8 rows
Straight from the federal item record. Entries the record leaves empty are shown empty — nothing is filled in on its behalf.
| MRC | Requirement | Value |
|---|---|---|
| ACDC | CURRENT TYPE | AC |
| ACZB | FREQUENCY RATING | HERTZ MINIMUM 50.0 AND HERTZ MAXIMUM 60.0 |
| AKYD | ACCESSORY COMPONENTS AND QUANTITY | ELECTRO-OPTICAL MAGNIFICATION READER 1; WORK STATION 1; VENOUS-BLOOD PIPETTER 1 |
| ALUE | UNIVERSAL MEDICAL DEVICE NOMENCLATURE SYSTEM CODE | 17741 |
| CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | ANALYZER, CLINICAL CHEMISTRY |
| ELEC | VOLTAGE IN VOLTS | 120.0 AND 220.0 |
| FEAT | SPECIAL FEATURES | MICROPROCESSOR CONTROLLED ELECTRO-OPTICAL INSTRUMENT FOR MEASURING PACKED CELL VOLUMES AND CONVERTING TO RELATIVE VALUES;COMPONENTS INC.75X MONOCULAR MICROSCOPE,ROTATING TUBE CARTRIDGE ASSY;OPICAL ENCODER;UV ILLUMINATOR LAMP;VARIOUS LENSES AND FILTERS TO CAUSE STAINED WHITE CELLS AND PLATELETS TO FLUORESCE;12 PLACE HEMATOCRIT CENTRIFUGE WITH HEAD COVER,FIXED FIVE MINUTE TIMER;12,000 RPM SPEED CAPACITY;AND RCF OF 14,387 X G;TWO POSITION PIPETTER WITH FIXED VOLUME OF 111.1 MICROLITERS;WORK STATION WITH NUMBERED SLOTS FOR CAPILLARY TUBES;WELLS FOR WHOLE BLOOD TUBES AND RACK FOR STORAGE OF PIPETTER;OPERATION METHOD:ELECTRIC;EXTERIOR CONTAINER QUANTITY:1;THE NUMBER OF ALTERNATING CURRENT PHASE:TWO |
| SUPP | SUPPLEMENTARY FEATURES | CONT C/O TWO SEALED CALIBRATION CHECK TUBES; ONE VENOUS,ONE CAPILLARY; ONE PAIR FORCEPS; SPARE PIPETTER "O" RINGS AND GREASE,SPARE LAMPS FUSES,TWO THRUST PIVOTS,ONE 5/64 IN. HEX KEY AND TEST GUIDE; CAPABLE OF PERFORMING THE FOLLOWING QUANTITATIVE FUNCTIONS ON A CENTRIFUGAL BLOOD TUBE; HEMOGLOBIN,HEMATOCRIT,PLATELET COUNT,WHITE BLOOD COUNT,GRANULOCYTE COUNT,AND LYMPHOCYTE-MONOCYTE COUNT; FOR CAPILLARY MICROHEMATOCRIT TUBE KIT ORDER 6630-01-233-7594; FOR VENOUS MICROHEMATOCRIT TUBE KIT,ORDER 6630-01-233-7593 AND VENOUS HEMATOCRIT TUBE KIT 6630-01-233-9592; REQUISITIONER SHOULD SPECIFY THE VOLTAGE AND FREQUENCY REQUIRED |
Cross-references
8 manufacturer part numbers on file for this NSN — each shown against the CAGE code it belongs to.
| Manufacturer part number | CAGE |
|---|---|
| 428605 | 22790 |
| 4451 QBCII | 22790 |
| 428605 | 4AZY5 |
| 6630NCM162618 | 64616 |
| 1397399 | 88739 |
| 6630-0284-0000 | 99998 |
| 4465 | 12717 |
| 4466 | 12717 |