NSN 6640-01-737-0713 · NIIN 017370713 · FSC 6640
NO DASHES 6640017370713
In Stock
THERMOCYCLER,LABORATORY
Laboratory Equipment and Supplies
Technical characteristics
2 rows
Straight from the federal item record. Entries the record leaves empty are shown empty — nothing is filled in on its behalf.
| MRC | Requirement | Value |
|---|---|---|
| FEAT | SPECIAL FEATURES | CONSISTS OF A 4-WELL HEATING BLOCK AND HEATING LID. THE TEMPERATURE IS PROGRAMMABLE BETWEEN 30 C AND 105 C WITH +/- 0.1 C PRECISION. THE RESISTIVE HEATING AND AIR FORCED COOING SYSTEMS ARE CONTROLLED BY A PID ALGORITHM FOR CYCLICAL HEATING/COOLING OF THE HEATING BLOCK. THE WATSON S HEATING RAMP RATE IS RATED AT 4 +/- 0.5 DEGREES/SEC. THE HEATING BLOCK IS COOLED BY 5 CFM (CUBIC FEET PER MINUTE) CAPACITY COOLING FAN PRODUCING 2 +/- 0.5 DEGREES/SEC COOLING RAMP TIME. THE HEATED LID PREVENTS WATER CONDENSATION DURING SAMPLE HEATING STEP. WATSON IS OPERATED BY A 4-POSITIONED DIP SWITCH. IT DOES NOT REQUIRE A COMPUTER OR MOBILE PHONE FOR OPERATION. STORE UP TO 16 DIFFERENT PCR PROGRAMS IN ITS FLASH MEMORY. THE MACHINE COMES WITH 16 DEFAULT PROGRAMS PRELOADED THAT CAN COVER ALM |
| CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | WATSON THERMOCYCLER, 4-WELL PORTABLE |