NSN 6665-00-161-8120 · NIIN 001618120 · FSC 6665
NO DASHES 6665001618120
In Stock
RADIACMETER
Hazard-Detecting Instruments and Apparatus
Technical characteristics
11 rows
Straight from the federal item record. Entries the record leaves empty are shown empty — nothing is filled in on its behalf.
| MRC | Requirement | Value |
|---|---|---|
| ABHP | OVERALL LENGTH | INCHES NOMINAL 8.000 |
| ABKW | OVERALL HEIGHT | INCHES NOMINAL 6.000 |
| ABMK | OVERALL WIDTH | INCHES NOMINAL 4.000 |
| AMKD | INDICATOR TYPE | METER |
| AMRM | RADIATION TYPE AND FUNCTION | BETA RAYS DETECTS AND GAMMA RAYS DETECTS AND X-RAYS DETECTS AND BETA RAYS MEASURES AND GAMMA RAYS MEASURES AND X-RAYS MEASURES |
| AMRX | DETECTING ELEMENT TYPE | IONIZATION CHAMBER |
| APCB | PORTABILITY | PORTABLE |
| APPJ | RADIATION RANGE PER HOUR | ROENTGENS 0.0 TO 5.0 |
| APPN | ACCURACY IN PERCENT | AT FULL SCALE 0.0 TO 10.0 |
| FEAT | SPECIAL FEATURES | WEIGHT 0.85 KG; PORTABLE SURVEY METER CONSISTING OF A PRESSURIZED IONIZATION CHAMBER, MICROPROCESSOR-BASSED WITH COMBINED ILLUMINATED ANALOG/DIGITAL LCD; UTILIZES AN INFRARED COMMUNICATOR FOR CALIBRATION AND ALLOWS THE INSTRUMENT TO BE USED AS A REMOTE DETECTOR; INCLUDES PROGRAMMABLE FLASH ALARM WHICH CAUSES THE DISPLAY TO PULSATE AT A RATE OF ONCE PER SECOND WHEN MEASURED DOSE EXCEEDS A PRESET LIMIT; IONIZATION CHAMBER FILLED TO A PRESSURE OF 6 ATMOSPHERES; CAN BE USED TO FREEZE THE MAX. EXPOSURE RATE ENCOUNTERED; 5 OPERATING RANGES, 0 TO 500 MICRO R/H, 0 TO 5 MR/H, 0 TO 50 MR/H, 0 TO 500 MR/H AND 0 TO 5 R/HOUR |
| MATT | MATERIAL | CASE PLASTIC |