NSN 6685-01-224-4282  ·  NIIN 012244282  ·  FSC 6685
NO DASHES 6685012244282
In Stock

CALIBRATOR,THERMOMETER

Pressure, Temperature, and Humidity Measuring and Controlling Instruments

Technical characteristics

14 rows

Straight from the federal item record. Entries the record leaves empty are shown empty — nothing is filled in on its behalf.

Technical characteristics from the federal item record
MRCRequirementValue
ABFYOVERALL DEPTHINCHES NOMINAL 26.250
ABKWOVERALL HEIGHTINCHES NOMINAL 45.000
ABMKOVERALL WIDTHINCHES NOMINAL 45.250
ACYNAC VOLTAGE RATINGVOLTS MINIMUM 208.0 AND VOLTS MAXIMUM 230.0
ACZBFREQUENCY RATINGHERTZ NOMINAL 60.0
ALGCMOUNTING CONFIGURATIONFREE STANDING CABINET ON CASTORS
BDHRCALIBRATION TYPEDEG CELSIUS AND DEG FAHRENHEIT
BDHSINSTRUMENT CALIBRATION RANGECOLD BATH -100 TO +70 DEG C (-148 TO +158 DEG F); HOT BATH +60 TO +300 DEG C (+140 TO +572 DEG F)
BDHWCALIBRATION ACCURACY RATING-100 TO -40 DEG C - PORM 5 DEG C; -40 TO +150 DEG C - PORM 1 DEG C; +150 TO +300 DEG C - PORM 5 DEG C
CTJSUNPACKAGED UNIT WEIGHTPOUNDS MAXIMUM 450.000
FAAZPHASESINGLE
FEATSPECIAL FEATURESC/O TWO TEST WELLS, EACH WELL IS 2.6"DIA X 11/3" DEEP (60 CUBIC INCHES) AND HAVE A 0.25" DEEP MERCURY TRAP. TRANSFER FLUIDS: COLD BATH - HALOCARBON 0.8; HOT BATH - DOW 710.
FTLDFUNCTIONAL DESCRIPTIONCONSISTS OF 2 INDEPENDENTLY OPERABLE BATH SYSTEMS. EACH BATH HAS ITS OWN TANK WITH A CONSTANT LEVEL TEST WELL & PUMP. EACH HAS ITS OWN TEMPERATURE CONTROLLER, PROBE, & HEATING AND COLLING SYSTEMS. SEPERATE FRONT PANEL POWER CONTROLS, LABELED HOT BATH AND COLD BATH, ALLOW INDEPENDENT POWER UP AND USE.
RDALREFERENCE DATA AND LITERATURECOMMERCIAL MANUAL

Cross-references

5 manufacturer part numbers on file for this NSN — each shown against the CAGE code it belongs to.

Manufacturer cross-reference part numbers and CAGE codes
Manufacturer part numberCAGE
MLEE82-117B30653
70134U744
A-A-5941958536
701364841
530964841