VERIFICATION KIT,RADIAC

Hazard-Detecting Instruments and Apparatus

Technical characteristics

5 rows

Straight from the federal item record. Entries the record leaves empty are shown empty — nothing is filled in on its behalf.

Technical characteristics from the federal item record
MRCRequirementValue
CXCYPART NAME ASSIGNED BY CONTROLLING AGENCYTEST VERIFICATION KIT
FEATSPECIAL FEATURES2 RADIOACTIVE SOURCES ARE INCLUDED, 8 UCI CS-137 CESIUM AND 30 NCI TH-232 THORIUM; THE TEST FIXTURE IS USED TO PROVIDE A REFERENCE GEOMETERY FOR THE CESIUM SOURCE; THE THORIUM SOURCE IS ATTACHED TO A PLASTIC HOUSING THAT ACTS AS A REFERENCE GEOMETERY; REFERENCE GEOMETERIES ARE SHOWN ON THE THORIUM SOURCE AND THE TEST SOURCE KIT INSIDE LID; NON-AVIATION CRITICAL SAFETY ITEM. TECHNICAL DATA REF TO 11H2-2-31
FTLDFUNCTIONAL DESCRIPTIONCONSISTS OF THE TS-100 TEST SOURCES AND GCF-200 (V2) TEST FIXTURE, IS USED TO VERIFY THE OPERATION OF THE ADM300A RADIOLOGICAL ASSESSMENT KITS; THEKITS CONSIST OF AN ADM300A SURVEY METER AND PROBES; IF THE DISPLAYED READINGIS WITHIN THE LISTED EXPECTED READING THE ADM300 OR PROBE IS WORKING AND READY FOR USE
RADCRADIOACTIVE CONTENTCONTAINS RADIOACTIVE MATERIAL
RADDRADIONUCLIDES DATAMICROCURIES CESIUM (55) CS-137 8.000