NSN 6632-01-332-7885 · NIIN 013327885 · FSC 6632
NO DASHES 6632013327885
In Stock
ANALYZER,BLOOD GAS
Medical Laboratory Analysis Instruments
Technical characteristics
14 rows
Straight from the federal item record. Entries the record leaves empty are shown empty — nothing is filled in on its behalf.
| MRC | Requirement | Value |
|---|---|---|
| ABRN | HEIGHT | INCHES 8.400 |
| ACDC | CURRENT TYPE | AC |
| ACZB | FREQUENCY RATING | HERTZ NOMINAL 50.0 OR HERTZ NOMINAL 60.0 |
| AQXY | TEST TYPE FOR WHICH DESIGNED | BLOOD GASES AND ELECTROLYTES; HCO3; HCT; BASE EXCESS AND O2 SAT |
| CBBL | FEATURES PROVIDED | DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION |
| CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | BLOOD GAS ANALYZER 120/230V 50/60 HZ AC: |
| CYFV | METHOD OF ANALYSIS | POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
| CYFY | ANALYSIS TEST/MEASUREMENT RANGE | PO2 0-600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MM/L; NA 100 TO 200 MM/L; CA 0.25 TO 4.25 MM/L; HCT 10 TO 75 PCT; HCO3 3.0 TO 60 MM/L; TCO2 3.0 TO 60 MM/L; BASE EXCESS -30 TO +30 MM/L; O2 SAT 30 TO 99 PCT |
| CYFZ | TEST SAMPLE/SPECIMEN SIZE | 0.5 ML |
| CYHM | TEST SAMPLE/SPECIMEN TYPE | WHOLE BLOOD |
| DPTH | DEPTH | INCHES 18.300 |
| ELEC | VOLTAGE IN VOLTS | 120.0 AND 230.0 |
| FEAT | SPECIAL FEATURES | UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING AND FLUSH SOLUTIONS AND GASES PLUS A WASTE CONTAINER FOR PROCESSED BLOOD, CALIBRATING, AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS, EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATION EVERY HOUR AND ONE-POINT CALIBRATION EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE,AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT. |
| WDTH | WIDTH | INCHES 9.100 |