Home / MS parts / NSN 6525-00-965-2501
NSN 6525-00-965-2501  ·  NIIN 009652501  ·  FSC 6525
NO DASHES 6525009652501
In Stock

X-RAY APPARATUS,RADIOGRAPHIC AND

Imaging Equipment and Supplies: Medical, Dental, Veterinary

Technical characteristics

30 rows

Straight from the federal item record. Entries the record leaves empty are shown empty — nothing is filled in on its behalf.

Technical characteristics from the federal item record
MRCRequirementValue
BNBBTABLE TYPETILTING
CCFNSCREEN TYPEW/SPOT-FILM
CLPGTUBE COOLING METHODAIR
CLPHTUBE MOUNTING TYPETUBE STAND AND TABLE
AQZFCONTROL TYPEVERTICAL
ARNFTUBE QUANTITY2
CLPRTUBE STAND MOUNTING TYPEOVERHEAD
CLPTTILT RANGE90 DEG TRENDELENBURG TO VERTICAL RANGE OF POSITIONS
SUPPSUPPLEMENTARY FEATURESREQUISITION SHOULD STATE FLOOR-LOADING CAPACITY,CEILING STRENGTH AND HEIGHT,WALL AND FLOOR CONSTRUCTION (CONCRETE,WOOD,ETC),ELECTRICAL SUPPLY AND CAPACITY,DIAGRAM OF ROOM INDICATING DIMENSIONS,COMPATABILITY W/EXISTING FACILITIES,AND WHETHER OR NOT CONTRACTOR INSTALLATION SERVICE IS REQUIRED
CMGNSCREEN LENGTHINCHES MINIMUM 12.000
CMGPSCREEN WIDTHINCHES MINIMUM 12.000
CMGQCASSETTE SIZE8 IN. BY 10 IN.,10 IN. BY 12 IN.
CXCYPART NAME ASSIGNED BY CONTROLLING AGENCYX-RAY APP RAD&FLUOR300MA 150PKV 2 TUBE 230V 60HZ AC W/ROT ANODE
ELECVOLTAGE IN VOLTS220.0
FAAZPHASESINGLE
FEATSPECIAL FEATURESFLOOR MOUNTED TABLE W/MOTOR DRIVEN ADJ TOP; DESIGNED TO USE W/MOTION PICTURE AND TELEVISION CAMERAS; REQUISITION SHOULD STATE WHETHER WALL-MOUNTED,HIGH VOLTAGE TAKE UP REEL IS REQUIRED
AAXXMOUNTING TYPESTATIONARY
ABFFFURNISHED ITEMSBUCKY DIAPHRAGM
CLQDTRANSFORMER RECTIFICATION METHODELECTRON TUBE
AKCVDRIVE TYPEMOTOR
AKYDACCESSORY COMPONENTS AND QUANTITYREMOTE CONTROLLED POWER DRIVEN BODY SECTIONAL DEVICE 1,COLLIMATOR 1,COMPRESSION DEVICE 1,FLUOROSCOPIC IMAGE INTENSIFIER W/22 CM DIA PHORPHOR 1,FOOTREST 1,ADJUSTABLE PATIENT GRASP HANDLE 2,TABLE TILT INDICATOR 1,PHOTO-TIMER RECEIVER IN SPOT FILM DEVICE 1,PHOTO-TIMER RECEIVER IN THE BUCKEY AND/OR CASSETTE CHANGER /OPTIONAL/ 1,RADIATION SHIELD 1,MYELOGRAPHIC PATIENT SUPPORT 1,SHOULDER SUPPORT 1
APGFDESIGN TYPESTATIONARY
ARQEELECTRIC POWER RATINGWATTS NOMINAL 40000.0
AWGHTUBE TYPEDOUBLE FOCUS, ROTATING ANODE RADIOGRAPHIC AND DOUBLE FOCUS, ROTATING ANODE FLUOROSCOPIC
FREQFREQUENCY IN HERTZ60.0
BPKZRECTIFICATION TYPEFULL-WAVE
CLQFTRANSFORMER OUTPUT CAPACITY300 MA OUTPUT AT 150 KV PEAK
CLPXRADIOGRAPHIC TIMER TYPEEXPOSURE
CLPYFLUOROSCOPIC TIMER TYPESYNCHRONOUS MOTOR
APHEOPERATION METHODAUTOMATIC